As mentioned, the AMS2750E is a standard for the heat treatment. The “AMS” name in the standard is an abbreviation of “Aerospace Materials Specifications”. The standard is published by SAE Aerospace, part of SAE International Group. The first version of the AMS2750 standard was published in 1980. Followed by revisions: revision A in 1987, B also in 1987, C in 1990 and D in 2005. The current revision AMS2750E was published in 2012.
The AMS2750 standard was initially developed to provide consistent specifications for heat treatment through the aerospace supply chain. The use of the standard is audited by PRI (Performance Review Institute) for the Nadcap (National Aerospace and Defense Contractors Accreditation Program). Prior to Nadcap, aerospace companies each audited their own suppliers, so there was a lot of redundancy and duplication of efforts. In 1990, the PRI was established to administer the Nadcap program.
According to the standard itself, the scope of the AMS2750E standard is the following:
"This specification covers pyrometric (high temperature) requirements for thermal processing equipment used for heat treatment. It covers temperature sensors, instrumentation, thermal processing equipment, system accuracy tests, and temperature uniformity surveys. These are necessary to ensure that parts or raw materials are heat treated in accordance with the applicable specification(s)."
In some industrial processes, it is relatively easy to measure and check the quality of the final product and judge if the product fulfills the requirements after the process is complete. You may be able to simply measure the end product and see if it is good or not.
In other processes where it is not possible/easy/practical to measure the quality of the final product you need to have a very tight control and documentation of the process conditions, in order to be sure that the final product is made according to the requirements.
It is easy to understand that heat treatment is a process where you need to have a very good control of the process in order to assure that you get the required end product, especially since the products are mostly used by the aerospace industry.
The AMS2750E is predominantly designed for the aerospace industries. But the same standards and processing techniques can be used within any industry which requires control of the thermal processing of raw materials and manufactured components, such as automotive, rail and manufacturing.
The CQI-9 is a similar set of requirements for heat treatment, mainly aimed
Let’s discuss Test Instruments (calibrators) and what AMS2750E says about them.
A traceable calibration of different levels of measurement instruments is obviously required. The higher level standards are typically calibrated in an external calibration laboratory. The process measurements are calibrated internally using “field test instruments”.
Metrological Traceability is often described as traceability pyramid, or as a traceability chain, see below:
Traceability pyramid:
Traceability chain:
AMS2750E also specifies the calibration period and accuracy requirements for the different levels of instruments, below is what is said about the secondary standard instrument and field test instrument:
Sometimes it is easier to look at a visual, so let's look at this required calibration accuracy graphically for “field test instrument” and “secondary standard instrument”. And as the Centigrade and Fahrenheit are different, below is a graph of both for your convenience:
The AMS2750E standard specifies different thermocouple types for different usage. Types B, R
Anyhow, the standard has the same accuracy specification regardless of the thermocouple type. This is a slightly strange requirement, as different thermocouples have much different sensitivities.
In practice, this means that a test field instrument (calibrator) normally has a specification for millivoltage, and when this mV accuracy is converted to temperature it means that the calibrator normally has different specifications for different thermocouple types. Some thermocouple types have
For example - a calibrator can have an electrical specification of 4 microvolts at 0 V. With a K type, this 4 µV equals a temperature of 0.1 °C (0.2 °F), but for a S type, this equals 0.7°C (1.3°F), and for a B type it equals almost 2°C (3.6 °F). Therefore, calibrators normally have very different accuracy specifications for different thermocouple types.
So the standard having the same accuracy regardless of the thermocouple type is a bit strange requirement.
To illustrate the different sensitivities of different thermocouple types, please see the graphics below. The graph shows what kind of thermovoltage (Emf) is generated in different temperature by different thermocouple types:
Let’s take a brief look at the contents of the AMS2750E standard and further discuss a few key points in the standard.
The AMS2750E standard starts with sections:
1. Scope
2. Applicable documents
Chapter 3 “Technical Requirements” of AMS2570E includes the following key sections. (These sections are discussed in more details in the next chapters):
3.1 Temperature sensors
3.2. Instrumentation
3.3. Thermal processing equipment
3.4. System Accuracy Tests (SAT)
3.5. Furnace Temperature Uniformity Survey (TUS)
3.6. Laboratory furnaces
3.7. Records
3.8. Rounding
The remaining sections are:
4. Quality assurance provisions
5. Preparation for delivery
6. Acknowledgement
7. Rejections
8. Notes
Section 3.1 discusses temperature sensors. Some key bullets from that section:
The AMS2750E standard specifies the thermocouple sensors to be used, as well as the sensor wire types.
The voltage to temperature conversion standard to be used (ASTM E 230 or other national standards).
Correction factors may be used to compensate for the errors found in calibration.
The temperature range for the sensors used.
Allowance to use wireless transmitters.
Contents of a sensor calibration certificate.
The max length of sensor wire/cable.
The max number of usage of thermocouples in different temperatures.
Types of thermocouple sensors to be used, the use for thermocouples (primary calibration, secondary calibration, sensor calibration, TUS, SAT, installation, load sensing), calibration period for thermocouples, and maximum permitted error.
Section 3.2 covers the instrumentation that the sensors are used with. This includes control, monitoring, recording, calibration, instrumentation, etc.
Instruments need to be traceably calibrated.
Minimum resolution/readability of test instruments (1 °F or 1 °C).
Specifications for electronic records.
Contents of calibration sticker:
Date, due date, performed by, any limitations
Contents of calibration record:
Instrument identification, make and model, standard(s) used, calibration method, required accuracy, as found and as left data of each calibration point, offset, found/left, sensitivity, statement of acceptance or rejection, any limitations or restrictions, calibration date, due date, performed by, calibration company, signature, quality, organization approval.
Section 3.3 discusses the furnace classification and the temperature uniformity requirements in each class. Going from class 1 having uniformity requirement of ±5°F / ±3 °C, to class 6 with ±50 °F / ±28 °C.
Section 3.4 discusses the system accuracy tests (SAT). The SAT is an on-site test where the whole measurement loop (
SAT shall be performed with a “field test instrument,” specified in the standard’s Table 3. SAT should be performed periodically or after any maintenance. SAT interval is based on equipment class and instrumentation type.
SAT test records shall include:
identification of sensor calibrated
ID of reference sensor
ID of test instrument
date and time
set points
readings of furnace under test
test instrument readings
test sensor correction factors
corrected test instrument reading
calculated system accuracy difference
an indication of acceptance or failure
who performed the test
signature
quality organization approval
Section 3.5 is about furnace temperature uniformity survey (TUS). The
There are many detailed specifications for
The TUS report shall include:
furnace identification
survey temperatures
sensor location and identification including detailed diagrams
time and temperature data from all sensors
correction factors for sensors in each temperature
as found and as left offsets
corrected/uncorrected readings of all TUS sensors at each temperature
testing company identification and signature
identification of the person who performed the survey
survey start date and time
survey end date and time
test instrumentation identification
identification of pass or fail
documentation of sensor failures (when applicable)
summary of corrected plus and minus TUS readings at each temperature after stabilization
quality organization approval